MIC Chinese Simplified Chinese Traditional English
Inquiry Basket Inquiry Basket (0) item | Recent Visit History | Join Audited Suppliers

Home » Product Directory » Instruments & Meters » Analysis Instrument » Medical Electric Equipment Safety Performance Integrated Testing Apparatus (DAZ-Y)

Medical Electric Equipment Safety Performance Integrated Testing Apparatus (DAZ-Y)

Medical Electric Equipment Safety Performance Integrated Testing Apparatus (DAZ-Y)
Medical Electric Equipment Safety Performance Integrated Testing Apparatus (DAZ-Y)

Product Description

The apparatus designed and manufactured according with GB9706.1-1995, whose technical capacities &indexes totally accord with GB9706.1-1995, IEC601-1: 1998. It can test earthing impedance, power, leakage current, dielectric intensity, remainder voltage, and test automatically in the setting order by PLC( earthing impedance- leakage current -power-remainder voltage-dielectric intensity). LCD's menus is controlled by touch key, and measuring data can be printed by the printer in the apparatus.

Technical data:
1. Earthing impedance test: Output current 1A, 10A, 25A, error: ± 3%± 2
Testing Range: 1~250m次, error: ± 3%± 2
Alarm figure: 0.05次, 0.1次, 0.15次, 0.2次, 0.5次, error: ± 3%± 2
2. Leakage current test:
Testing Range: 0.5米A~10mA, error: ± 5%, MD impedance, DC-1MHz Virtual Value.
Alarm figure: 0.01mA, 0.05mA, 0.1mA, 0.5mA, 1mA, 2.5mA, 5mA, 10mA, error± 5%.
Voltage of power supply: AC 121V, 242V error: ± 2%, Output Power: 2KW (if 242V)
Voltage of adding press: AC 121V, 242V, 250V error: ± 2%, Output Power: 100W(250V)
3. Power test:
Single-phase virtual power: 50~2000W(if 242V), error: ± 3%± 2
Output voltage: 99V, 110V, 121V, 198V, 220V, 242V, error: ± 3%± 2
4. Leakage current test:
Voltage of power supply: AC 121V, 242V error: ± 3%± 2
Testing Range: 3~60V error ± 1V
Give an alarm if output impedance >100M次, >60V
5. Dielectric intensity test:
Output voltage: 500V~5KV, error: ± 3%± 2, become the set figure in 10s
Alarm figure: 2mA, 5mA, 10mA, 15mA, error: ± 3%± 2

More Product Features

   
Trademark:  HY
Model NO.:  DAZ-Y
Standard:  IEC601-1:1988
Company: Shanghai Hengyi Medicine Making Equipment Co., Ltd.

Click here to send inquiry

Most Popular

Quick Products