Basic Info.
Ray Type
Gamma Ray Spectrometer
Spectrometer Detector
SDD
Stationary Phase
Solid Adsorbent
Chromatography Type
Gas Chromatography
Chromatographic Column Type
Capilary Column
Chromatographic Column Length
10~30cm
Chromatographic Column Diameter
20~40mm
Transport Package
Carton Packing
Product Description
Chemical Analysis Machine Gas Chromatography for Residual Solvents Testing
Technical index
Column Oven |
Size: | 28 × 30 × 18 cm |
Operating temperature range: | ambient temperature ~ 450ºC |
-Using liquid nitrogen cold trap: | - 80ºC ~ 400ºC |
-Using dry ice cold trap: | - 55ºC ~ 400ºC |
Temperature setpoint: | 0.1ºC |
Maximum heating speed: | 120 ºC/min |
Maximum run time: | 999.99 min |
Maximum Programmed temperature order: | 7 order |
Executable column loss compensation (dual channels) |
Heating area |
Six automatic heating zones control (Not including furnace temperature controlling system, two inlets, two detectors and two auxiliary heating zone) |
Maximum temperature of auxiliary heating zone : 300 ºC |
Separated columns oven | The temperature for each column oven can be set separately |
Inlets |
Dual channel Inlets |
Optional inlets: |
PPIP (can connect up a big caliber capillary column) |
SSI |
Cool On-Column |
PPI |
High precision electronic pressure/flow control |
Maximum service temperature: | 400 ºC |
Column capital pressure setting range: | 0 ~ 100 psi |
Column capital pressure control precision: | 0.01 psi |
Total flow setting range: | 0 ~ 100 mL/min |
Flow setting precision: | 0.1 mL/min |
Reducer union can fit in 1/4 inch, 1/8 inch packed column and 0.53/0.32 mm capillary column |
Split /Splitless Capillary Inlets |
High precision electronic pressure/flow control |
Maximum service temperature: | 400 ºC |
Column capital pressure setting range: | 0 ~ 100 psi (or 150psi) |
Column capital pressure control precision: | 0.01 psi |
Total flow setting range: | 0 to 1000 ml/min (helium gas) 0 to 200 ml/min (nitrogen gas) |
Flow setting precision: | 0.1 mL/min |
Maximum split ratio: | 1:1000 |
Cool On-Column |
High precision electronic pressure/flow control |
Column capital pressure control precision: | 0.01 psi |
Column capital pressure setting range: | 0 ~ 100 psi |
Maximum service temperature: | 400 ºC |
Furnace temperature tracking mode |
Self-editing temperature programmed mode |
Fit 0. 53/0. 32 mm capillary column |
Volatile concatenated inlet |
High precision electronic pressure/flow control |
Column capital pressure setting range: | 0 ~ 100 psi |
Column capital pressure control precision: | 0.01 psi |
Total flow setting range: | 0 ~ 1000 ml/min (helium gas or hydrogen gas ) |
Flow setting precision: | 0.1 ml/min |
Maximum service temperature: | 400 ºC |
Detector (Electronic flow control module fits for all detectors) |
Optional detectors: | Hydrogen flame ionization detector (FID) |
Thermal conductivity detector (TCD) |
Electron capture detector (Micro-ECD) |
Helium phosphorus detector (NPD) |
Flame photometric detector (FPD) |
Hydrogen flame ionization detector (FID) |
High precision electronic flow/pressure control |
Fit for packed column and capillary column |
Maximum service temperature: | 450 ºC |
Minimum LOD: | < 2.5 pg c/sec (n - C16) |
Linear dynamic range: | 107 (±l0 %) |
Data acquisition frequency: | Maximum 100 Hz |
Thermal Conductivity Detector (TCD) |
High precision electronic flow/pressure control |
Fit for packed column and capillary column |
Monofilament micro pool design |
Maximum service temperature: | 400 ºC |
Data acquisition frequency: | maximum100 Hz |
Linear dynamic range: | 105 (±5%) |
Minimum LOD: | < 400 pg propane /ml with Helium carrier gas |
Electron capture detector (ECD) |
High precision electronic flow/pressure control |
Fit for packed column and capillary column |
Maximum service temperature: | 400 ºC |
Detector complemental air type: | 5% methane/argon gas or nitrogen gas |
Data acquisition frequency: | Maximum 100 Hz |
Linear dynamic range: | >5 × 10^5 |
Minimum LOD: | < 0.01 pg/sec lindane |
Radioactive source: | < 12 mCi 63Ni |
Nitrogen Phosphorus Detector (NPD) |
High precision electronic flow/pressure control |
Fit for packed column and capillary column |
Maximum service temperature: | 450ºC |
Minimum LOD: | < 3 pg C/sec |
Linear dynamic range: | 105 N, 105 P |
Data acquisition frequency: | Maximum 100 Hz |
Minimum LOD: | < 0. 2 pg N/sec, < 0. 2 pg P/sec |
Selectivity: | 25000:1 gN/gC, 75000:1 gP/gC |
Flame Photometric Detector (FPD) |
High precision electronic flow/pressure control |
Maximum service temperature: 250ºC |
Minimum LOD: | < 4 pg S/sec, < 1 pg P/sec |
Selectivity: | 105 gS/gC, 106 gP/gC |
Linear dynamic range: | > 10³ S, 104 P |
Auxiliary EPC module |
3 channel auxiliary pressure control module (pressure control mode) |
2 channel auxiliary pressure control module (pressure control mode) |
Single channel auxiliary pressure control module (pressure control mode) |
Single channel gas flow control module (pressure/flow control mode) |
Autosampler |
8-bit autosampler Fit for Agilent 7683 liquid autosampler |
Data communications |
Ethernet (LAN), series interface RS - 232 Remote start/stop |
Instrument control, monitor, Method Editor, storage and call, they all can be run in the workstation |
Address:
Business Department, Building 2, No. 98 Wuyi Road, Changsha, Hunan, China
Business Type:
Manufacturer/Factory, Trading Company
Business Range:
Chemicals, Industrial Equipment & Components, Instruments & Meters
Management System Certification:
ISO 9001
Company Introduction:
Hunan Xiangnan Chemicals Import&Export Co. Ltd has been dedicated to producing and constantly improving the quality of our products since then. Our company main products are Gas Chromatography (GC), Mass Spectrometry (MS), Gas Mass Spectrometry (Gas-MS) and Atomic Absorption Spectrometer(AAS) and Atomic Fluorescence Spectrometer(AFS), MSM(Methyl Sulfonyl Methane)and DMSO(Dimethyl Sulfoxide) we own the modern standard workshops and production equipment and we are the leading manufacturer in China with a product range for many applicatiions, With years of export experience along with excellent quality, advanced services and competitive prices. HNXN has won numerous customers′trust and support.